TMHX-C series - General purpose infrared thermometers TMHX-C.pdf (PDF)TMHX-C
Very cost-effective and the ideal choice for installing in assembly lines and first-time customer.
TMZ9

FEATURES

Measuring range 0°C to 500°C

Cost-Effective
The TMHX-CNE0500(E)series can be configured equipment you have already had. That makes them cost-effective.

Stability remarkably impreved
The TMHX series using indium antmonide(InSb) remarkably imprpves the stability of measurement compared to the previos model,the TMH91,which uses a thermopile element.

A wide variety of accessories
We have a wide variety of accessories that canbe accurately measured in adverse environments.

Displays and setting units
We have various displays and parameter setting units for setting the tempreature indications of the infrared thermometers.


Easy alignment with "the optical LED sihgt"
It is critical to accurately focus on a target in order to get a precise measurement.
The THMH series uses “the optical coaxial LED sight”that can emit the visible light to sight on target for easy alignment.


SPECIFICATIONS/MEASURING DISTANCE AND TARGET-SIZE
Type [1]TMHX-CNE0500-0035E1.6
[2]TMHX-CNE0500-0070E003
[3]TMHX-CNE0500-0120E5.5
Measuring range 0 to 500°C
Wave length 2 to 6.8μm
Detection element InSb
Measuring distance [1]35mm
[2]70mm
[3]120mm
Target-size [1]φ1.6mm
[2]φ3.0mm
[3]φ5.5mm
Sight LED Sight
Accuracy Under 300°C ; ±3.0°C
Upper300°C ; Measured value±1.0%
Repeatability ± 0.5°C
Measuring resolution Less than 0.5°C
Response time 0.01 to 5 s(0 to 95% analog output)
Warm-up time 1 min
Output Analog output(Non-isolation)
Output type(Changeable)
DC4 to 20mA,DC0 to 20mA,mV/°C,DC0 to 1V
(Option : 0 to 5V / 0 to 10V)
Output RS232C
output(RS232C confirmity,non-isolation)
Output swing range : ±4V, Communicating speed
: 4800,9600,19200,38400,57600,115200 bps
Alarm output (Non-isolation) Open drain : DC27V, 0.2A
Peak Hold Reset method (Selection)@[1] Time : 0.01 to 10 s changeable@
[2]Discharge : Time 0.01 to 10 sec, level 0.2 to 1.0
Emissivity correction Guaranteed range ; 0.3 to 1.0 with reflection correction
Setting range ; 0.050 to 1.000(Setting resolution 0.001)
Internal VR ; Built-in
Sensor correction function Offset ; -50 to +50°C(°F)
Supply voltage DC4.7 to 27V, 0.1A Max
Analog output 0 to 5V ; 8 to 27 V/ 0 to 10 V ; 13 to 27V
Cable length(standard) 2m
Dimensions of each parts HX-C1