L-2000 is a two colour infrared thermometer designed to measure temperature under sever surrounding conditions. The thermometer employs combination of two sensors, i,e,, lInGaAs and Silicon sensors, that exhibit different sensitivity in the same wavelength. Thus the thermometer can measure two colours simultaneously.
By using this combination of sensors, it is possible to measure temperature of small objects through dust smoke, fumes and through glass windows in the optical path.
SPECIFICATIONS OF L-2000
Temperature range
600 - 2,000°C
Element
InGaAs, Si (Two colour)
Wavelength
1.6µm, 1.9µm
Accuracy rating
±1% ε=1.0
Repeatability
±1% ε=1.0
Resolution
1°C
Measured value display
7segLED Character quantity 14.6mm
Preset value display
7segLED Character quantity 10.1mm
Response time
0.05 sec.
Compensattion range
0.050 - 1.200
Setting
Key operation on front panel
Analog output
DC0 - 10V, 4-20mA load resistance 470Ω or low
Alarm output
Open collector output Hi, Lo (standard)
Max DC30V 50 mA
Relay output Hi, Lo (option)
Contact arrangement 1a, AC 250V8A, DC30V5A